Strip Test Database
STRIP21
Tests
- DC characterization
- Bandpass characterization
- Bandpass characterization
- Long acquisition
- Y-factor test
- Y-factor test
- OMT short acquisition with HEMT's turned off
- RF test -60dBm no crossguide signal injected at 0°(rectangular waveguide horizontal)
- RF test -60dBm no crossguide signal injected at 0°(rectangular waveguide horizontal)
- RF test -60dBm no crossguide signal injected at 0° default
- RF test -60dBm no crossguide signal injected at 45°
- RF test -60dBm no crossguide signal injected at 45°
- RF test -60dBm no crossguide signal injected at 45°
- RF -60dBm no cross guide signal injected at 90°
- RF -60dBm no cross guide signal injected at 90°
- RF -60dBm no cross guide signal injected at 90°
- OMT short acquisition with HEMT's turned off
- RF test -60dBm no crossguide signal injected at 22.5°
- RF test -60dBm no crossguide signal injected at 22.5°
- RF test -60dBm no crossguide signal injected at 22.5°
- 300K and 77K in front of circular feed horn
- 300K and 77K in front of the circular feed horn
- 300K and 77K in front of circular feed horn
- polarization grid tilted 45°. 77K in front of the feed horn and 300 at 90°
- polarization grid tilted 45°. 77K in front of the feed horn and 300 at 90°
- RF -60dBm no cross guide signal injected at 67.5°
- RF -60dBm no cross guide signal injected at 67.5°
- RF -60dBm no cross guide signal injected at 67.5°
- OMT Y-factor 77-300K in front of circular feed horn
- OMT Y-factor 77-300K in front of circular feed horn
- OMT long acquisition
Bandpass estimates
- Bandpass characterization : νcen = 43.3±0.7 GHz, Δν = 6.5±0.8 GHz (details, analysis report).
- Bandpass characterization : νcen = 43.5±0.8 GHz, Δν = 6.7±1.2 GHz (details, analysis report).
Noise temperature estimates
- Y-factor test : Tnoise = 92.7±16.8 K (details, analysis report).
- Y-factor test : Tnoise = 117.2±24.5 K (details, analysis report).
Spectrum estimates
- Long acquisition : fknee = ± Hz, wn = ± ADU2/√Hz (details, analysis report).