Strip Test Database
STRIP37
Tests
- DC characterization
- DC characterization - Tambient
- Bandpass characterization Tambient (PS0101 - PWR3 blind)
- Bandpass characterization Tambient (PS0110 - PWR0 blind)
- Y-factor, warm-up
- Y-factor, cool-down
- DC characterization
- Bandpass characterization
- Bandpass characterization
- Short acquisition with HEMTs turned off
- Long acquisition
Bandpass estimates
- Bandpass characterization : νcen = 43.3±0.2 GHz, Δν = 6.5±0.2 GHz (details, analysis report).
- Bandpass characterization : νcen = 43.5±0.3 GHz, Δν = 6.7±0.5 GHz (details, analysis report).
Noise temperature estimates
- Y-factor, warm-up : Tnoise = 38.9±6.4 K (details, analysis report).
- Y-factor, cool-down : Tnoise = 29.7±4.8 K (details, analysis report).
Spectrum estimates
- Long acquisition : fknee = ± Hz, wn = ± ADU2/√Hz (details, analysis report).