Strip Test Database
STRIP64
Tests
- DC characterization - Tamb
- Short acquisition with HEMTs turned off
- Short acquisition with HEMTs turned off
- Bandpass characterization (PWR3 blind)
- Bandpass characterization (PWR0 blind)
- Y-factor, warm-up (PWR0 blind)
- Y-factor, cool-down (PWR3 blind)
- Long acquisition
- DC characterization - Tcryo - repeated
- Bandpass characterization Tcryo repeated (PS0101 - PWR3 blind)
- Bandpass characterization Tcryo repeated (PS0110 - PWR0 blind)
- Y-Factor test_repeated_up LB (PS0101-PWR0 blind)
- Y-Factor test_repeated_down LB (PS0110-PWR3 blind)
- Long acquisition repeated
- Short acquisition with HEMTs turned off - repeated
- Short acquisition - repeated - cooler turned-off - PS 0101
- Short acquisition - repeated - cooler turned-off - PS switching
Bandpass estimates
- Bandpass characterization (PWR3 blind) : νcen = 43.5±0.2 GHz, Δν = 8.8±0.0 GHz (details, analysis report).
- Bandpass characterization (PWR0 blind) : νcen = 43.2±0.1 GHz, Δν = 7.7±0.4 GHz (details, analysis report).
- Bandpass characterization Tcryo repeated (PS0101 - PWR3 blind) : νcen = 43.1±0.1 GHz, Δν = 8.0±0.1 GHz (details, analysis report).
- Bandpass characterization Tcryo repeated (PS0110 - PWR0 blind) : νcen = 43.1±0.9 GHz, Δν = 7.7±1.9 GHz (details, analysis report).
Noise temperature estimates
- Y-factor, warm-up (PWR0 blind) : Tnoise = 29.1±2.3 K (details, analysis report).
- Y-factor, cool-down (PWR3 blind) : Tnoise = 31.1±2.4 K (details, analysis report).
Spectrum estimates
- Long acquisition : fknee = ± Hz, wn = ± ADU2/√Hz (details, analysis report).
- Long acquisition repeated : fknee = ± Hz, wn = ± ADU2/√Hz (details, analysis report).