Strip Test Database

STRIP31 – Short acquisition with HEMTs turned off

General information

Edit  Delete

Polarimeter STRIP31
Acquisition date
BandQ
Cryogenic?Yes
OperatorsSilvia Caprioli, Federico Incardona
PHSW stateN/A
Notes

**These are the offsets to use for Y-factor computation**

test taken at noon.
Output values are much higher with respect to the other short acquisition with HEMTs turned off, because of a thermal drift.

The source data file was «SCI170421_122216 - STRIP31_Steady300s_P=0_21Apr2017_bis.txt» (410 kB)

Authorsilvia
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1812 1866 1816 1845 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
10050 9050 5970 8860 Edit  Delete

Reports

Noise temperatures
Bandpasses
Spectrum