Strip Test Database

STRIP52 – Long acquisition

General information

Edit  Delete

Polarimeter STRIP52
Acquisition date
BandQ
Cryogenic?Yes
OperatorsClaudio Pincella
PHSW stateswitching
Notes

The source data file was «SCI171109_133319 - STRIP52_1_f__PS-default_24hours_09Nov2017.txt» (128429 kB)

Authorclaudio
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1905 1802 1846 1860 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
81840 80130 79000 79970 Edit  Delete
HEMT biases
HEMT Vd [V] Id [mA] Vg [mV]
H0 (HA1) 0.9 5.0 -37.0
H1 (HB1) 0.89 2.56 -91.0
H2 (HA2) 1.2 15.27 208.0
H3 (HB2) 1.2 13.45 190.0
H4 (HA3) 0.9 10.0 100.0
H5 (HB3) 0.92 9.01 299.0
Temperatures
TA1 [K] TA2 [K] TB1 [K] TB2 [K] Tp1 [K] Tp2 [K] Tc1 [K] Tc2 [K]
20.0 25.6 20.0 27.3 20.0 16.6 15.0 25.5 Edit  Delete

Reports

Noise temperatures
Bandpasses
Spectrum
Method αI αQ fkQ [mHz] WNQ [mK2/Hz] αU fkU [mHz] WNU [mK2/Hz] Details Report Actions
0.97700±0.00000 0.834±0.001 410.0±70.0 1.30±0.200 0.359±0.000 70.0±20.0 1.10±0.100 JSON Download Edit  Delete