Strip Test Database

STRIP30 – Long acquisition

General information

Edit  Delete

Polarimeter STRIP30
Acquisition date
BandQ
Cryogenic?No
OperatorsClaudio Pincella
PHSW stateswitching
Notes

The source data file was «SCI170927_154740 - STRIP30_1_f__PS-default_24hrs_27Sept2017.txt» (100366 kB)

Authorclaudio
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1881 1820 1862 1883 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
82250 84150 80895 83210 Edit  Delete
HEMT biases
HEMT Vd [V] Id [mA] Vg [mV]
H0 (HA1) 0.9 11.7 -180.0
H1 (HB1) 0.9 9.3 -105.0
H2 (HA2) 0.9 11.0 222.0
H3 (HB2) 0.9 11.0 326.0
H4 (HA3) 0.9 11.0 -98.0
H5 (HB3) 0.9 4.0 154.0
Temperatures
TA1 [K] TA2 [K] TB1 [K] TB2 [K] Tp1 [K] Tp2 [K] Tc1 [K] Tc2 [K]
20.0 25.8 20.0 27.9 20.0 16.0 10.0 16.5 Edit  Delete

Reports

Noise temperatures
Bandpasses
Spectrum
Method αI αQ fkQ [mHz] WNQ [mK2/Hz] αU fkU [mHz] WNU [mK2/Hz] Details Report Actions
0.73598±0.00000 0.474±0.000 5700.0±2300.0 3.00±0.600 0.491±0.000 6100.0±2600.0 1.90±0.400 JSON Download Edit  Delete