Strip Test Database

STRIP58 – Long acquisition

General information

Edit  Delete

Polarimeter STRIP58
Acquisition date
BandQ
Cryogenic?Yes
OperatorsSilvia Caprioli, Federico Incardona
PHSW stateswitching
Notes

The source data file was «SCI171102_151644 - STRIP58_1_f__PS-default_24hours_02Nov2017.txt» (147681 kB)

Authorsilvia
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1887 1936 1895 1893 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
79550 82300 80400 79000 Edit  Delete
HEMT biases
HEMT Vd [V] Id [mA] Vg [mV]
H0 (HA1) 0.9 3.0 175.0
H1 (HB1) 0.9 4.03 53.0
H2 (HA2) 1.2 12.2 272.0
H3 (HB2) 1.21 14.0 164.0
H4 (HA3) 0.91 9.02 316.0
H5 (HB3) 0.9 9.07 434.0
Temperatures
TA1 [K] TA2 [K] TB1 [K] TB2 [K] Tp1 [K] Tp2 [K] Tc1 [K] Tc2 [K]
20.0 25.8 20.0 27.5 20.0 17.2 15.0 25.4 Edit  Delete

Reports

Noise temperatures
Bandpasses
Spectrum
Method αI αQ fkQ [mHz] WNQ [mK2/Hz] αU fkU [mHz] WNU [mK2/Hz] Details Report Actions
1.11815±0.00000 1.410±0.020 130.0±80.0 0.90±0.100 1.220±0.020 20.0±10.0 0.74±0.090 JSON Download Edit  Delete