Strip Test Database

STRIP34 – Long acquisition

General information

Edit  Delete

Polarimeter STRIP34
Acquisition date
BandQ
Cryogenic?Yes
OperatorsClaudio Pincella
PHSW stateswitching
Notes

The source data file was «SCI171219_154324 - STRIP34_1_f__PS-default_24hours_19Dec2017.txt» (160041 kB)

Authorclaudio
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1840 1821 1816 1874 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
97230 97220 96350 93400 Edit  Delete
HEMT biases
HEMT Vd [V] Id [mA] Vg [mV]
H0 (HA1) 0.9 8.61 117.0
H1 (HB1) 0.9 5.67 0.0
H2 (HA2) 0.9 9.76 504.0
H3 (HB2) 0.9 5.16 119.0
H4 (HA3) 0.9 11.9 157.0
H5 (HB3) 0.9 4.3 0.0
Temperatures
TA1 [K] TA2 [K] TB1 [K] TB2 [K] Tp1 [K] Tp2 [K] Tc1 [K] Tc2 [K]
20.0 25.0 20.0 27.2 20.0 16.5 15.0 25.8 Edit  Delete

Reports

Noise temperatures
Bandpasses
Spectrum
Method αI αQ fkQ [mHz] WNQ [mK2/Hz] αU fkU [mHz] WNU [mK2/Hz] Details Report Actions
0.67688±0.00000 1.328±0.001 20.0±4.0 1.80±0.400 0.967±0.001 18.0±4.0 1.60±0.400 JSON Download Edit  Delete