Strip Test Database

STRIP13 – Long acquisition

General information

Edit  Delete

Polarimeter STRIP13
Acquisition date
BandQ
Cryogenic?Yes
OperatorsClaudio Pincella
PHSW stateswitching
Notes

The source data file was «SCI180227_165515 - STRIP13_1_f__PS-default_polV2_24hours_27Feb2018.txt» (144718 kB)

Authorclaudio
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1874 1816 1796 1796 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
100760 102330 101265 104135 Edit  Delete
HEMT biases
HEMT Vd [V] Id [mA] Vg [mV]
H0 (HA1) 1.0 5.18 -233.0
H1 (HB1) 1.0 3.62 579.0
H2 (HA2) 0.9 6.31 506.0
H3 (HB2) 0.9 14.11 460.0
H4 (HA3) 0.3 20.7 289.0
H5 (HB3) 0.9 17.43 368.0
Temperatures
TA1 [K] TA2 [K] TB1 [K] TB2 [K] Tp1 [K] Tp2 [K] Tc1 [K] Tc2 [K]
20.0 25.6 20.0 27.9 20.0 15.8 15.0 25.3 Edit  Delete

Reports

Noise temperatures
Bandpasses
Spectrum
Method αI αQ fkQ [mHz] WNQ [mK2/Hz] αU fkU [mHz] WNU [mK2/Hz] Details Report Actions
0.71201±0.00000 2.516±0.008 14.0±2.0 2.50±0.300 2.076±0.004 9.4±0.8 2.20±0.300 JSON Download Edit  Delete