Strip Test Database

STRIP77 – Bandpass characterization - PS 0101

General information

Edit  Delete

Polarimeter STRIP77
Acquisition date
BandW
Cryogenic?Yes
OperatorsClaudio Pincella
DescriptionBandpass characterization - PS 0101
PHSW state0101
Notes

The source data file was «SCI180427_105733 - Wband60_Tcryo_RF+5dBm_0101_doppio-attenuatore_27Apr2018.txt» (847 kB)

Vey bad balance

Authorf.benedini
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1877 1972 1869 1985 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
103700 101350 101670 101200 Edit  Delete
HEMT biases
HEMT Vd [V] Id [mA] Vg [mV]
H0 (HA1) 0.403 4.79 150.0
H1 (HB1) 0.496 5.09 231.0
H2 (HA2) 0.401 5.29 122.0
H3 (HB2) 0.544 4.6 91.0
H4 (HA3) 0.401 5.29 267.0
H5 (HB3) 0.544 4.6 215.0
Temperatures
TA1 [K] TA2 [K] TB1 [K] TB2 [K] Tp1 [K] Tp2 [K] Tc1 [K] Tc2 [K]
20.0 17.1 20.0 15.6 23.0 28.4 24.1 20.0 Edit  Delete

Reports

Noise temperatures
Bandpasses
Method Central frequency Bandwidth Details Report Actions
98.70±0.36 8.69±0.54 JSON Download Edit  Delete
Spectrum