Strip Test Database

STRIP37 – Y-factor, cool-down

General information

Edit  Delete

Polarimeter STRIP37
Acquisition date
BandQ
Cryogenic?Yes
OperatorsClaudio Pincella
DescriptionY-factor, cool-down
PHSW state0110
Notes

The source data file was «SCI171205_120200 - STRIP37_Y-Factor_LBdown_50-40-30-20_PS0110_Pol-V4_05Dec2017.txt» (1889 kB)

Authorclaudio
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1800 1820 1812 1819 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
101200 99500 97150 100100 Edit  Delete
HEMT biases
HEMT Vd [V] Id [mA] Vg [mV]
H0 (HA1) 0.9 6.15 32.0
H1 (HB1) 0.9 8.0 51.0
H2 (HA2) 0.9 4.89 303.0
H3 (HB2) 0.9 7.06 418.0
H4 (HA3) 0.9 6.0 -141.0
H5 (HB3) 0.9 9.0 92.0
Temperatures
TA1 [K] TA2 [K] TB1 [K] TB2 [K] Tp1 [K] Tp2 [K] Tc1 [K] Tc2 [K]
20.0 24.9 50.0 59.3 20.0 17.2 15.0 25.7 Edit  Delete
20.0 24.9 40.0 48.1 20.0 17.2 15.0 25.7 Edit  Delete
20.0 24.9 30.0 36.1 20.0 17.2 15.0 25.7 Edit  Delete
20.0 24.9 20.0 26.7 20.0 17.2 15.0 25.7 Edit  Delete

Reports

Noise temperatures
Method Tnoise [K] GQ1 [ADU/K] GU1 [ADU/K] GU2 [ADU/K] GQ2 [ADU/K] Details Report Action
non-linear fit 29.7±4.8 2020±228 2655±224 2016±169 1606±219 JSON Download Edit  Delete
Bandpasses
Spectrum