Strip Test Database

STRIP73 – Bandpass characterization - STRIP case2 - PS1001

General information

Edit  Delete

Polarimeter STRIP73
Acquisition date
BandW
Cryogenic?Yes
OperatorsFederica Benedini, Claudio Pincella
DescriptionBandpass characterization - STRIP case2 - PS1001
PHSW state1001
Notes

The source data file was «SCI180424_132017 - Wband31_Tcryo_RF-5dBm_1001_caso2_Vd_700mV_H0_H1_+50mV_H2_H3_24Apr2018.txt» (892 kB)

Authorclaudio
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1932 1989 1993 1933 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
100430 98220 97900 98300 Edit  Delete
HEMT biases
HEMT Vd [V] Id [mA] Vg [mV]
H0 (HA1) 0.7 4.66 334.0
H1 (HB1) 0.7 7.79 153.0
H2 (HA2) 0.37 6.25 280.0
H3 (HB2) 0.405 5.21 210.0
H4 (HA3) 0.37 6.25 280.0
H5 (HB3) 0.405 5.21 210.0
Temperatures
TA1 [K] TA2 [K] TB1 [K] TB2 [K] Tp1 [K] Tp2 [K] Tc1 [K] Tc2 [K]
20.0 17.1 20.0 15.9 23.0 28.8 24.0 20.0 Edit  Delete

Reports

Noise temperatures
Bandpasses
Method Central frequency Bandwidth Details Report Actions
99.33±0.43 8.92±1.22 JSON Download Edit  Delete
Spectrum