Strip Test Database

STRIP64 – Short acquisition - repeated - cooler turned-off - PS switching

General information

Edit  Delete

Polarimeter STRIP64
Acquisition date
BandQ
Cryogenic?Yes
OperatorsClaudio Pincella, Mario Zannoni
DescriptionShort acquisition - repeated - cooler turned-off - PS switching
PHSW stateswitching
Notes

The source data file was «SCI180508_141633 - prova_senza_cooler_default.txt» (548 kB)

The test is aimed to check the possible effect of the cooler operation on the high noise level on the detector output.

The conclusion is that the cooler cycle has no effect.

To be noted the saturation of some detector output due to the increasing temperature of the polarimeter

Authorclaudio
Date of creation
JSON recordLink
Detector outputs

Reports

Noise temperatures
Bandpasses
Spectrum