Strip Test Database

STRIP28 – Y-Factor test_up LB (PS1001-PWR3 blind)

General information

Edit  Delete

Polarimeter STRIP28
Acquisition date
BandQ
Cryogenic?Yes
OperatorsClaudio Pincella
DescriptionY-Factor test_up LB (PS1001-PWR3 blind)
PHSW state1001
Notes

The source data file was «SCI171130_172200 - STRIP28_Y-Factor_LBup_20-30-40_PS1001_Pol-STRIP++_30Nov2017.txt» (1348 kB)

Authorclaudio
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1775 1795 1809 1762 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
83400 84800 84600 95300 Edit  Delete
HEMT biases
HEMT Vd [V] Id [mA] Vg [mV]
H0 (HA1) 0.9 6.5 -292.0
H1 (HB1) 0.9 6.5 -332.0
H2 (HA2) 0.9 5.17 259.0
H3 (HB2) 0.9 3.9 197.0
H4 (HA3) 0.9 11.67 420.0
H5 (HB3) 0.9 12.75 429.0
Temperatures
TA1 [K] TA2 [K] TB1 [K] TB2 [K] Tp1 [K] Tp2 [K] Tc1 [K] Tc2 [K]
20.0 25.6 20.0 27.2 20.0 16.7 15.0 25.6 Edit  Delete
20.0 25.6 30.0 36.4 20.0 16.7 15.0 25.6 Edit  Delete
20.0 25.6 40.0 48.4 20.0 16.7 15.0 25.6 Edit  Delete

Reports

Noise temperatures
Method Tnoise [K] GQ1 [ADU/K] GU1 [ADU/K] GU2 [ADU/K] GQ2 [ADU/K] Details Report Action
non-linear fit 40.1±3.9 698±50 944±57 725±44 587±44 JSON Download Edit  Delete
Bandpasses
Spectrum