Strip Test Database

STRIP72 – Y-factor - PS 1001 down

General information

Edit  Delete

Polarimeter STRIP72
Acquisition date
BandW
Cryogenic?Yes
OperatorsFederica Benedini
DescriptionY-factor - PS 1001 down
PHSW state1001
Notes

The source data file was «SCI180525_112225 - STRIP_W25_Y-Factor_LB__down_50-40-30-20K_PS1001_25May2018.txt» (2371 kB)

Authorf.benedini
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1886 1913 1964 1888 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
98130 93630 96800 97210 Edit  Delete
HEMT biases
HEMT Vd [V] Id [mA] Vg [mV]
H0 (HA1) 0.5 4.86 413.0
H1 (HB1) 0.5 3.4 376.0
H2 (HA2) 0.44 5.44 342.0
H3 (HB2) 0.4 5.88 356.0
H4 (HA3) 0.44 5.44 342.0
H5 (HB3) 0.4 5.88 356.0
Temperatures
TA1 [K] TA2 [K] TB1 [K] TB2 [K] Tp1 [K] Tp2 [K] Tc1 [K] Tc2 [K]
20.0 19.8 50.0 40.3 24.0 19.8 20.0 17.4 Edit  Delete
20.0 19.8 40.0 32.9 24.0 19.8 20.0 17.3 Edit  Delete
20.0 19.8 30.0 25.4 24.0 19.9 20.0 17.3 Edit  Delete
20.0 19.7 20.0 15.4 24.0 19.8 20.0 17.3 Edit  Delete

Reports

Noise temperatures
Method Tnoise [K] GQ1 [ADU/K] GU1 [ADU/K] GU2 [ADU/K] GQ2 [ADU/K] Details Report Action
non-linear fit 84.7±9.3 374±36 488±41 520±43 279±30 JSON Download Edit  Delete
Bandpasses
Spectrum