Strip Test Database

STRIP75 – Y-factor - PS 1001 down

General information

Edit  Delete

Polarimeter STRIP75
Acquisition date
BandW
Cryogenic?Yes
OperatorsFederica Benedini
DescriptionY-factor - PS 1001 down
PHSW state1001
Notes

The source data file was «SCI180530_102940 - STRIP_W46_Y-Factor_LB__down_50-40-30-20K_PS1001_30May2018.txt» (2065 kB)

Authorf.benedini
Date of creation
JSON recordLink
ADC offsets
PWR0 PWR1 PWR2 PWR3
1960 2023 2104 2002 Edit  Delete
Detector outputs
PWR0 PWR1 PWR2 PWR3
112370 104250 104520 105760 Edit  Delete
HEMT biases
HEMT Vd [V] Id [mA] Vg [mV]
H0 (HA1) 0.499 4.96 73.0
H1 (HB1) 0.504 5.24 199.0
H2 (HA2) 0.411 6.19 532.0
H3 (HB2) 0.412 4.26 496.0
H4 (HA3) 0.411 6.19 532.0
H5 (HB3) 0.412 4.26 496.0
Temperatures
TA1 [K] TA2 [K] TB1 [K] TB2 [K] Tp1 [K] Tp2 [K] Tc1 [K] Tc2 [K]
20.0 16.6 50.0 38.9 23.0 18.4 20.5 17.5 Edit  Delete
20.0 16.5 40.0 31.1 23.0 18.4 20.5 17.5 Edit  Delete
20.0 16.5 30.0 23.4 23.0 18.3 20.5 17.5 Edit  Delete
20.0 16.4 20.0 15.1 23.0 18.3 20.5 17.5 Edit  Delete

Reports

Noise temperatures
Method Tnoise [K] GQ1 [ADU/K] GU1 [ADU/K] GU2 [ADU/K] GQ2 [ADU/K] Details Report Action
non-linear fit 70.7±4.4 819±43 1045±48 1255±56 659±39 JSON Download Edit  Delete
Bandpasses
Spectrum