Strip Test Database

STRIP25 – rectangular feed horn 77K and 300K in front of the horn

General information

Edit  Delete

Polarimeter STRIP25
Acquisition date
BandQ
Cryogenic?Yes
OperatorsMario Zannoni
Descriptionrectangular feed horn 77K and 300K in front of the horn
PHSW state0110
Notes

The source data file was «SCI180717_130828 - STRIP25_OMT_0110_rectan_horn_300_77K_congrid_vert_e senza_17jul2018.txt» (304 kB)

in this test we put 77K in front of feed horn. the we put the polarization grid between the load and ther horn with vertical wires that should reflect in the horn a mix of polarimeter noise and surrounding ambient temperature.
Successively with the grid in the same position we put 300K load behind the grid and than we remove the grid

Authorluoni
Date of creation
JSON recordLink
Detector outputs

Reports

Noise temperatures
Bandpasses
Spectrum